Durability and Failure Types of S2-alar-iliac Screws: A Finite Element Analysis Study

Seung-Jae Hyun, Ki-Jeong Kim
Journal of Advanced Spine Surgery 2020;10(2):33-38.
Published online: December 31, 2020
Department of Neurosurgery, Spine Center, Seoul National University Bundang Hospital, Seoul National University College of Medicine, Seongnam, Korea
Corresponding author:  Seung-Jae Hyun, Tel: +82-31-787-7169, Fax: +82-31-787-4097, 
Email: hyunsj@snu.ac.kr
  • 13 Views
  • 2 Download
  • 0 Crossref
  • 0 Scopus
next

Background
S2-alar-iliac (S2AI) screws are one of the options for spinopelvic fixation to improve stability across the lumbosacral junction. The S2AI screws cross the cortical surfaces of the sacroiliac joint, which can increase the biomechanical strength of the instrumentation. Objective: To investigate the durability and failure types of S2AI screw by finite element model (FEM) analysis.
Methods
Through the FEM, complex material and geometrical properties of the biological system can be evaluated, and various physical variables, such as stress, and fracture, can be analyzed. We examined the biomechanical stress distribution at the set screw and screw head by using a FEM. Von Mises (V.M.) stress (MPa) is derived from 3-dimensional status of stress. The finite element software Abaqus® version 6.5 (ABAQUS Inc., Johnston, RI, USA) was used to create a FEM.
Results
We quantified the peak V.M. stress applied to the set screw and screw head when rod to S2AI screw trajectory angle was 30º angled and perpendicular. In FEM analysis, at an angle of 30 degrees rather than perpendicular, the stress increased further around the area where the screw head and rod contacted and the displacement distribution of set screw also increased.
Conclusion
S2AI screw fixation has several drawbacks such as screw fracture and dislodgement of the set screw. This FEM analysis can support the result.

Download Citation

Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:

Include:

Durability and Failure Types of S2-alar-iliac Screws: A Finite Element Analysis Study
J Adv Spine Surg. 2020;10(2):33-38.   Published online December 31, 2020
Download Citation
Download a citation file in RIS format that can be imported by all major citation management software, including EndNote, ProCite, RefWorks, and Reference Manager.

Format:
  • RIS — For EndNote, ProCite, RefWorks, and most other reference management software
  • BibTeX — For JabRef, BibDesk, and other BibTeX-specific software
Include:
  • Citation for the content below
Durability and Failure Types of S2-alar-iliac Screws: A Finite Element Analysis Study
J Adv Spine Surg. 2020;10(2):33-38.   Published online December 31, 2020
Close
TOP